Patient doses from screen-film and full-field digital mammography in a population-based screening programme

Author(s)

Publication date

2011-02-17

Series/Report no

Radiation protection dosimetry;

Publisher

Oxford University Press

Document type

Abstract

The aim of this study was to compare mean glandular dose (MGD) in all full-field digital mammography (FFDM) and screen film mammography (SFM) systems used in a national mammography screening program. MGD from 31 screening units (7 FFDM and 24 SFM), based on an average of 50 women at each screening unit, representing 12 X-ray models (6 FFDM and 6 SFM) from five different manufacturers were calculated. The MGD was significantly lower for FFDM compared with SFM (craniocaudal): 1.19 versus 1.27 mGy, respectively, mediolateral oblique: 1.33 versus 1.45 mGy, respectively), but not all of the FFDM units provided lower doses than the SFM units. Comparing FFDMs, the photon counting scanning-slit technology provides significantly lower MGDs than direct and indirect conversion digital technology. The choice of target/filter combination influences the MGD, and has to be optimised with regard to breast thickness.

Keywords

Version

This is a pre-copy-editing, author-produced PDF of an article accepted for publication in Radiation protection dosimetry following peer review. The definitive publisher-authenticated version Hauge, I. H. R., Pedersen, K., Sanderud, A., Hofvind, S. & Olerud, H. M. (2011). Patient doses from screen-film and full-field digital mammography in a population-based screening programme. Radiation protection dosimetry is available online at: http://dx.doi.org/10.1093/rpd/ncq598.